CloudTesting™ Lab 2.32.00

CloudTesting™ Lab Rev 2.32.00

1. Summary of revision

  • Revision number Rev 2.32.00
  • Reason for revision
    Addition of function, Fixing issue

2. Addition of functions

2.1 Data Log

(1) The function to change lengthes of data log fields
The number of characters to output data log in text format can be specified.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

For details of the above function, refer to the following section of the user’s manual.
“Controlling a Data Log”

2.2 Main Window

(1) Function to open Work Project folder
The work project folder can be opened from the button on the toolbar.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(2) Function to refer to Pattern Programming Manual
The Pattern Programming Manual can be referred to from the buttons on the Main Window.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

For details of the above functions, refer to the following section of the user’s manual.
“Using Main Window”

2.3 CT Command

(1) Function to prohibit editing the work project
The following option have been added to CT_Start.bat.

  • -noedit
    Prohibits editing the work project.
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

For details of the above function, refer to the following section of the user’s manual.
“Details of CT Commands which Control CloudTesting(TM) Lab startup”

2.4 Expert Mode

(1) The function to setting of pre-processing and post-processing for analysis tools
The function to settings of pre-processing and post-processing for analysis tools.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

For details of the above function, refer to the following section of the user’s manual.
“Analysis Tools” in the “Menu Bar” section of the “Expert Mode” chapter

3. Fixed issues

3.1 Setup and execution

(1) Memory leak occurred in the settings of measurement condition
Memory leak occurred in some settings of the following measurement conditions:

  • CX1000_MCU

    • DC parametric measurement condition
    • ADC DC Linearity measurement condition
    • Pattern condition
  • CX1000_MEMORY_TYPE_F)

    • Fail capture condition
    • Signal condition

Behavior after improvement
The above memory leak does not occur.

Revisions in which this phenomenon occursTarget firmware
Rev 2.20.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

(2) Memory leak occurred in specific IPs
Memory leak occurred in the following IPs:

  • Occurred in executing

    • System Condition Monitor Tool (PCXT01-M6020)
    • System Condition Monitor Tool for Memory (PCXT01-M8003)
    • Measurement facilitation by General I/F(I2C) package IP (PCXA01-M1029)
    • Operating Power Supply Voltage Search IP (PCXA01-M5040)
  • Occurred in setting the conditions)

    • DC Parametric Measure IP (PCXA01-M1025)
    • Power Supply Current Measure IP (PCXA01-M1028)
    • Frequency Measure IP (PCXA01-M1090)

Behavior after improvement
The above memory leak does not occur.

Revisions in which this phenomenon occursTarget firmware
Rev 2.20.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

(3) The expression set to the register value of the pattern condition was cleared
When the value in pattern object and the value set by evaluating specified expression were same, the expression was cleared.

Behavior after improvement
The expressions are not cleared.

Revisions in which this phenomenon occursTarget firmware
Rev 2.22.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

3.2 Message Display Area

(1) The variable values were not applied to the logging file name
The variable values could not be applied to the logging file name containing variables when runtime error had occurred.

Behavior after improvement
Even if runtime error has occurred, the variable values are applied to the logging file name containing variables.

Revisions in which this phenomenon occursTarget firmware
Rev 2.30.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

(2) Collapsed log was shown
The log containing link text could collapse.

Behavior after improvement
The correct log is shown.

Revisions in which this phenomenon occursTarget firmware
Rev 1.00.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F, CX50_GENERAL

(3) Link text was not saved to the logging file
Link text was not saved to the logging file.

Behavior after improvement
Link text is saved to the logging file.

Revisions in which this phenomenon occursTarget firmware
Rev 1.00.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F, CX50_GENERAL

3.3 Work Project

(1) An unexpected confirmation dialog box was displayed
An unexpected confirmation dialog box was displayed before the display of error dialog box when the work project for Expert Mode was opened in Standard Mode from the Startup Window.

Behavior after improvement
The unexpected confirmation dialog box is not displayed.

Revisions in which this phenomenon occursTarget firmware
Rev 2.27.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

3.4 Expert Mode

(1) An error occurred on copying a flow item
When satisfying the following all conditions, an error occurred on copying a flow item. * [Power off After Measurement] setting existed * The measurement algorithm was changed

Behavior after improvement
The above error does not occur.

Revisions in which this phenomenon occursTarget firmware
Rev 2.20.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

(2) The flow item could not be set to the [Target Measure Item] properly in the settings for preprocessing and/or post-processing when the analysis tool was started
In the following case, the flow item could not be set to the [Target Measure Item] properly in the settings for preprocessing and/or post-processing when the analysis tool was started. * The flow items selected before starting the analysis tool contain the flow items that cannot be specified as the [Target Measure Item].

Behavior after improvement
The flow item is set to the [Target Measure Item] properly.

Revisions in which this phenomenon occursTarget firmware
Rev 2.27.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F

3.5 License

(1) Returning license process colud not complete
When returning license process had failed for communication failure, re-authentication was required on same computer.

Behavior after improvement
In situation above, the license is available until completing returning license process on the computer.

Revisions in which this phenomenon occursTarget firmware
Rev 1.00.00 ~ Rev 2.31.03CX1000_MCU, CX1000CX1000_MEMORY_TYPE_F, CX50_GENERAL