CloudTesting™ Lab 2.27.00

1. Summary of revision

Revision number
Rev 2.27.00
Reason for revision
Addition of function, Functional improvement, Fixing issue

2. Addition of functions

2.1 Flow Execution (Standard Mode)

(1) The measurement items can be grouped

The multiple measurement items can be grouped. Each group can be selected to execute.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

2.2 IP Manager

(1) Creating a log file

IP Manager creates a log file.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

2.3 Expert Mode

(1) Algorithm names are additionally saved into list of flow items

Algorithm names are additionally saved into list of flow items.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(2) Power-off control in list view of flow items

You can edit power-off control in list view of flow items.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

3. Improved functions

3.1 Work Project

(1) Improvement of the error handling for the work project name

Improve error handling when a work project name has invalid characters.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

3.2 Setup and Execution

(1) Displaying [TS] name

Add [TS] name to [TS] display field in signal condition tab.

Target firmware
CX1000_MCU

3.3 CT Command

(1) Improvement of execution time

Shorten the overhead time of the execution.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

3.4 Pattern Compile Window

(1) Improvement of the error handling

Improve error handling when a pattern program can not be saved to a folder.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

3.5 IP Manager

(1) Downloading the proper revision of IPs

The revision of the downloaded IPs are the latest one for the using CloudTesting Lab.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

(2) Selection for using Expert Mode License

You can select to use Expert Mode License.

(Notes on update installation) If all of the following conditions are satisfied, the Expert Mode License is deselected and the startup window is started without Expert Mode.

  • IPs to be used are selected in IP Manager settings
  • When installing, the installed IPs and IP Manager settings are not deleted

For using the Expert Mode License, select it in the IP Manager Window.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(3) Improvement of communication via network proxy

Improve to communicate via network proxy.

Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

3.6 Expert Mode

(1) Change the initialization rule of Control Word

The initialization rule of Control Word is changed as follows:

  • Execute measurement items in flow: Initialize(Unchanged)
  • Execute measurement items to be selected: Do not initialize(Changed)
  • Execute a measurement item in Analysis Tool: Do not initialize(Unchanged)
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4. Fixed issues

4.1 Work Project

(1) ‘#’ cannot to be included in work project name

‘#’ could not be included in work project name.

Behavior after improvement
‘#’ can be included in work project name.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.2 Setup and Execution

(1) Error dose not occer in the certain pattern execution

No error ocurred in the following conditions.

  1. The JSR branch destination written in main pattern does not exists in the subroutine patterns.
  2. Even though the main pattern refers to a subroutine, the subroutine pattern is not specified.
Behavior after improvement
In the conditions above, an error occur.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.26.00
Target firmware
CX1000_MCU

(2) Can not be executed SCAN pattern after the executing DSRC pattern

SCAN pattern could not be executed after the executing DSRC pattern.

Behavior after improvement
In the conditions above, SCAN pattern can be executed correctly.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.26.00
Target firmware
CX1000_MCU

4.3 Analysis Tools

(1) Change the launch process of analysis tools

An analysis tool can select unexecutable measurement item.

Behavior after improvement
An analysis tool can not select unexecutable measurement item.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.4 CT Command

(1) Variable can not be deleted

The variable which is set or retrieved by CT Command could not be deleted.

Behavior after improvement
The variable can be deleted.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(2) CT_VariableGet.bat - An array variable of string type cannot be displayed correctly

If an array variable of string type includes empty string element, the empty string could not be displayed in the list.

Behavior after improvement
An array variable of string type includes empty string can be displayed in the list correctly.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(3) CT_WorkProjectOpen.bat cause an error

An error occurred by executing CT_WorkProjectOpen.bat repeatedly.

Behavior after improvement
An error do not occur by executing CT_WorkProjectOpen.bat repeatedly.
Revisions in which this phenomenon occurs
Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.5 IP Manager

(1) The license authentication is unavailable

If CloudTesting ™ Lab was forcibly terminated, authentication of CloudTesting ™ Lab at launch was unsuccessful until Windows restarted.

Behavior after improvement
License authentication is performed correctly.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

4.6 Expert Mode

(1) Cloud Testing™ Lab hang-up

Cloud Testing™ Lab hung-up on selecting [Execute Select Item & Suspend] menu after editing a signal condition.

Behavior after improvement
In the condition above, Cloud Testing™ Lab dose not hang-up.
Revisions in which this phenomenon occurs
Rev 2.22.00 to Rev 2.26.00
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F