CloudTesting™ Lab 2.25.00

1. Summary of revision

Revision number
Rev 2.25.00
Reason for revision
Addition of function, Functional improvement, Fixing issue

2. Addition of functions

2.1 CT Command

(1) LoggingMode option for CT_DataLogSet.bat was supported

LoggingMode option for CT_DataLogSet.bat was supported.

Target firmware
CX1000_MCU

3. Improved functions

3.1 Setup and Execution

(1) The driver wave drawn in Signal Condition was improved

The driver wave drawn in Signal Condition was imporved as below : - It is drawn with dashed line before the level is settled, and it is drawn with solid line after the level settled. - It is drawn with gray line before the driver is enabled, and it is drawn with red line after the driver is enabled.

Target firmware
CX1000_MCU

3.2 Analysis Tools

(1) Performance improvement for loading pattern comments

Performance for loading pattern comments was improved.

Target firmware
CX1000_MCU

4. Fixed issues

4.1 CT Command

(1) The measurement item name containing some spaces could not be specified as a command argument

The measurement item name containing some spaces could not be specified as a command argument for the following CT commands:

  • CT_FlowStart.bat
  • CT_FlowStartNoWait.bat
  • CT_MeasrureItemStart.bat
  • CT_MeasureItemStartNoWait.bat
  • CT_MeasureItemSelect.bat
  • CT_FlowItemStart.bat
  • CT_FlowItemStartNoWait.bat
Behavior after improvement
The double-quoted measurement item name containing some spaces can be specified.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(2) CT Command could be unusable if the Operation Wondow for CX1000 was running

CT Command could be unusable if the Operation Wondow for CX1000 was running.

Behavior after improvement
CT Command can be used properly.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.2 Variable

(1) An error occured when the array of strings was renamed

An error occurred when the array of strings was renamed if the initial values for the array elements were specified by index numbers.

Behavior after improvement
The variable name can be renamed properly.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F
(2) The variable which definition had not been completed was displayed on
the Select Variable
dialog box in the Expert Mode
The variable which definition had not been completed was displayed on
the Select Variable
dialog box which was used to select the variable for the Branch Switch Value
or the interpose processing in the Expert Mode.
Behavior after improvement
The variable which definition has not been completed is not displayed. The measurement cannot be started if there is such variable on the Define Variables
dialog box.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.3 Analysis Tools

(1) The datalog was set disabled by interrupting the analysis tool execution

The datalog was set disabled by interrupting the analysis tool execution.

Behavior after improvement
The datalog is not set disabled in the situation above.
Revisions in which this phenomenon occurs
Rev 2.22.00 to Rev 2.24.01
Target firmware
CX1000_MCU

(2) If there was the measurement item whose name was ended with “shmoo“, Shmoo could not be execute

If the measurement item named “XXX” and the measurement item named “XXX_shmoo” existed, the Shmoo for “XXX” could not be executed in the Expert Mode. (XXX is an arbitrary string)

Behavior after improvement
Shmoo works properly in the situation above.
Revisions in which this phenomenon occurs
Rev 2.22.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

(3) Analysis tools could not be stopped executing by stop button

Analysis tools could not be stopped executing by stop button.

Behavior after improvement
The above behavior is improved.
Revisions in which this phenomenon occurs
Rev 1.00.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F

4.4 Flow Execution (Standard Mode)

(1) The work project which had been saved after deleting some measurement items could not be opened

The work project which had been saved after deleting the measurement item that had been selected in the Setup and Execution
could not be opened.
Behavior after improvement
The saved work project can be opened properly.
Revisions in which this phenomenon occurs
Rev 2.24.00 to Rev 2.24.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F