CloudTesting™ Lab 2.21.02

Reason for revision

Functional improvement

Functional improvement

IP Manager

Network connection could not be established through proxy.

There was a case where the proxy setting of Windows Internet option was not applied to the IP Manager. Therefore, the communication error occurred at the starting of CloudTesting™ Lab.

The conditions of occurrence are as follows: - One of the checkboxes is set on the [Automatic configuration] area in the [Local Area Network(LAN) Settings].

Behavior after improvement
Proxy setting of Windows Internet option is applied to the IP Manager under the above conditions. And if network connection still can not be established, CloudTesting™ Lab supports IP installer package as backup.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.21.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

Setup and Execution

A slight memory leak occurred while executing the measure items.

A slight memory leak (several hundred bytes) occurred while executing the measure items.

Behavior after improvement
The memory leak was improved.
Revisions in which this phenomenon occurs
Rev 2.20.00 to Rev 2.21.01
Target firmware
CX1000_MCU, CX1000_MEMORY_TYPE_F, CX50_GENERAL

When a experiment item was executed continuously for a long time,

the available memory became insufficient. When a experiment item in an academic package was executed continuously for a long time, the data displayed on the meter gradually increased. As a result, the available memory became insufficient.

Behavior after improvement
Upper limit of the number of data to be displayed on the meter was limited to the latest 100,000. As a result, the available memory in long continuous execution is no longer insufficient.
Revisions in which this phenomenon occurs
Rev 2.21.00 to Rev 2.21.01
Target firmware
CX50_GENERAL