Reason for revision
Functional improvement
Improvement items
Setup and Execution
The PowerSupply does not output voltage and current correctly.
When “Source Range” is selected “-1 to +7” in the setting source volutage conditions, specified current was not output. For more detail, please see below.
https://store.cts-advantest.com/download/UpdateNotice/CloudTestingLab_Rev2.23.01_Update_note_en.pdf
- Behavior after improvement
- The PowerSupply outputs voltage and current correctly.
- Phenomenon occurs revision
- Rev 1.00.00 ~ Rev 2.23.00
- Target firmware
- CX1000_MCU, CX1000_MEMORY_TYPE_F