How to solve a runtime exception of "The protection circuit worked.".

Question

The following error is popped up after executed work-project.
What the root cause for that.

Answer

This error occures that the PBVCC current exceeds 500mA in executing work-project.
Please confirm the following points.

  • Check that the amount of current in the connected circuit to PBVCC does not exceed 500mA.
  • Check that PBVCC is not shorted.
Updated at