What condition is the I/O channel when the DC parametric measurement is executed?

Question

The DC parametric measurement unit is able to interrupt the I/O channels and measure the voltage or current of the I/O channels.
What condition is the I/O channel when the DC parametric measurement is executed?

Answer

In DC Parametric Measure IP, the measurement is executed in the following sequence.
(1)Turn off the measurement target channel.
(2)Detach the I/O channel path from the measurement target channel.
(3)Connect the DC parametric measurement unit to the measurement target channel.
(4)Measure.
(5)Detach the DC parametric measurement unit from the measurement target channel.
(6)Connect the I/O channel path to the measurement target channel.
(7)Turn on the measurement target channel.