What is required the operating condition of the CloudTesting ™ station?

Question

What is required the operating condition of the CloudTesting ™ station?

Answer

The following conditions must be required.

1.When used with the accuracy of DC calibration (light)

  • Within 1 year since the internal standard of Cloud Testing™ Station is calibrated
  • Ambient temperature is in the range of +20°C to +30°C
  • Within 24 hours since DC calibration (light) is executed

2.When used with the accuracy of DC calibration (fine)

  • Within 1 year since the internal standard of Cloud Testing™ Station is calibrated
  • Ambient temperature is in the range of +20°C to +30°C, and the fluctuation of the ambient temperature is within ±1°C
  • Within 24 hours since DC calibration (fine) is executed

For the installation environment of CloudTesting ™ Station, refer to “Safety Information” and “System Specifications” in “CX1000 Users Manual”.

For details on the accuracy of CloudTesting(TM) Station, refer to “System Specification” in “CX1000 User’s Manual”.

For details on the traceability of CloudTesting(TM) Station, refer to “Traceability” in “CX1000 User’s Manual”.

Calibration expiration date of CloudTesting ™ Station internal standard can be confirmed with a sticker on top side of CloudTesting ™ Station casing.

For DC Calibration (Light) execution procedure, refer to “Calibrating This Instrument” in “CX1000 User’s Manual”.

For DC Calibration (Fine) execution procedure, refer to “Executing Precise Level Calibration” in “CX1000 User’s Manual”.